Differential Pulse Anodic Voltammetric Determination of Chlorzoxazone in Pharmaceutical Formulation using Carbon Paste Electrode

Authors

  • Sayed I. M Zayed Faculty of Industrial Education, Beni Suef University
  • Yousry M Issa Chemistry Department, Faculty of Science, Cairo University

DOI:

https://doi.org/10.17344/acsi.2020.5731

Keywords:

Chlorzoxazone, Carbon paste electrodes, Differential pulse anodic voltammetry Pharmaceutical dosage form, Human urine samples.

Abstract

The electrochemical behavior of chlorzoxazone at the carbon paste electrode was investigated in 0.04 mol/L Britton-Robinson buffer pH 6.50 using cyclic and differential pulse voltammetric techniques. Cyclic voltammetric studies indicated that the oxidation of the drug was irreversible and controlled mainly by diffusion. Experimental and instrumental parameters were optimized (50 mV/s scan rate, 50 mV pulse amplitude, and 0.04 mol/L Britton-Robinson (BR) buffer pH 6.50 as a supporting electrolyte) and a sensitive differential pulse anodic voltammetric method has been developed for the determination of the drug over the concentration range 0.17-1.68 µg/mL chlorzoxazone, with detection and quantitation limits of 0.05 and 0.16 µg/mL, respectively. The proposed voltammetric method was successfully applied to the determination of the drug in its pharmaceutical formulation (Myoflex tablets), and in spiked human urine samples.

Published

15.12.2020

Issue

Section

Analytical chemistry